A 68-year-old patient with symptomatic, decompensated heart failure (NYHA functional class III, NT-proBNP: 18151 ng/l) secondary to dilated/non-ischemic cardiomyopathy was admitted for decongestion, re-rhythmization, and implantation of a CRT-D system. The patient had a history of long persistent atrial fibrillation and heart failure with severely reduced left ventricular function despite more than three months of optimal medical therapy (OMT). In addition, a left bundle branch block (LBBB) with a QRS duration of 185 ms and moderate mitral regurgitation were present.
Three months prior, coronary artery disease had been excluded by coronary angiography, ruling out an ischemic etiology of the underlying cardiomyopathy.
The initial ECG demonstrated atrial fibrillation with a ventricular heartrate of approximately 70 bpm and LBBB with a QRS duration of 185 ms (Fig. 1).

Figure 1 Speed: 25mm/s, Amplitude: 5 mm/mV Baseline 12-lead ECG showing atrial fibrillation with a ventricular rate of 70 bpm, LBBB, and a QRS duration of 185 ms.
Figure 2 Baseline chest X-ray

Significant clinical improvement was achieved through negative fluid balance by applying intravenous diuretic therapy.
After transesophageal-echocardiographic exclusion of intracardiac thrombi, the patient was successfully cardioverted to sinus rhythm. Amiodarone was subsequently initiated for antiarrhythmic therapy.
Echocardiography revealed severely reduced left ventricular ejection fraction (LVEF) of 25%, with global hypokinesia to akinesia, as well as marked left ventricular dilatation (LVEDD 64
mm, EDV 212 ml), and significantly enlarged left atrium (LAESVI 66 mL/m²). In addition, moderate secondary mitral regurgitation with a central jet was observed (EROA 0.38 cm², regurgitant volume 58 ml, VC 6 mm).

Figure 5 Bulls Eye – Speckle Trecking, EF 25%

Implantation procedure
On-site equipment: Ziehm Vision RFD C-arm and Amedtec 12-lead ECG system
Following local anesthesia, an approximately 4-cm skin incision was made in the left infraclavicular region. After identification and preparation of the cephalic vein, the vessel was opened using a venotomy. The atrial and ventricular leads were then introduced through the cephalic vein under fluoroscopic guidance. Both leads were positioned and screwed into the myocardium (RV lead: Plexa MRI S 65, Biotronik; RA lead: Solia S 53, Biotronik).
Following this, a CS guiding catheter (Selectra Extended Hook 45, Biotronik) was introduced through a 9-F sheath, and the coronary sinus (CS) was visualized using contrast. Cannulation of the CS proved challenging due to the presence of a venous valve (Fig. 6 and 7). Despite the use of various coronary guidewires, repeated attempts to cross the valve and gain access to the CS were unsuccessful.

Due to the inability to implant a conventional CRT-D system, the strategy was changed to left bundle branch pacing (LBBP).
The axillary vein was punctured extrathoracically. A Selectra 3D 65-42 catheter (Biotronik) was advanced through a 9-F sheath and positioned in the left bundle branch area under contrast imaging and fluoroscopic angulation. Due to inadequate lead fixation, the lead had to be repositioned several times. Ultimately, successful implantation of the Solia CSP S 60 lead (Biotronik) was achieved using pace mapping, with favorable electrical results.

Electrical parameters:
QRS duration: 125 ms
Stimulus-to-end of QRS: 140 ms
R-wave peak time (RWPT): 70 ms
V6–V1 interpeak interval: 45 ms

RA-lead: sensing 7.2 mV, pacing threshold 1.0 V / 0.4 ms, impedance 537 Ω
RV-lead: sensing 12.3 mV, pacing threshold 0.3 V / 0.4 ms, impedance 713 Ω, shock impedance 59 Ω
LV-lead: sensing 24.2 mV, pacing threshold 0.4 V / 1.0 ms, impedance 752 Ω

The leads were connected to a Rivacor 3 HF-T CRT-D system (Biotronik). The entire system was placed into the previously prepared device pocket, together with a gentamicin sponge, and secured with a fixation suture. The procedure was completed with continuous layered subcutaneous closure followed by a resorbable intradermal suture.
DDDR biventricular pacing was programmed, with an RV offset of 80 ms relative to the LV channel to minimize right ventricular pacing. The detection zones were programmed to VT 188 bpm, VF 222 bpm, and a monitor zone at 171 bpm.
The postoperative course was uneventful. A posteroanterior (PA) and lateral chest X-ray demonstrated appropriate lead positioning and excluded pneumothorax (Fig. 8 and 9).

Postoperatively, the characteristic transition from non-selective to selective left bundle branch pacing was observed during pacing threshold testing at 0.9 V / 1.0 ms (Fig. 10).

Figure 10 Speed: 25 mm/s, Amplitude 5 mm/mv. Transition from non-selective to selective left bundle branch pacing. Blue arrow: non-selective left bundle branch pacing. Green arrow: selective left bundle branch pacing.

The patient was discharged in an improved general condition.
At the follow-up visit two months later, the wound was fully healed without signs of irritation or infection, and the device function was normal. Of particular note was the significant clinical improvement in the patient’s heart failure symptoms (NYHA functional class I-II, NT-proBNP of 3256 ng/l and HFimpEF of 36%) and also an increased functional capacity and ability to perform activities of daily living, which in this case was extensive gardening.
In this case, the initial decision to pursue conventional cardiac resynchronization therapy was based on the presence of dilated/non-ischemic cardiomyopathy with an LVEF of 25%, LBBB with a QRS duration of 185 ms, and more than three months of optimal medical therapy (OMT). However, in this case LBB pacing proved to be the only available and effective approach for correction of the LBBB.