The 76-year-old male patient was admitted to our Cardiology Department in January 2023 for pacemaker implantation due to 2:1 and intermittent third-degree AV-block (Figure 1).

Considering the expected high right ventricular pacing burden, conduction system pacing was planned. Preoperative transthoracic echocardiography revealed normal systolic left ventricular ejection fraction without regional wall motion abnormalities and significant valvular disease.
During the procedure, the atrial lead was temporarily positioned in the right ventricle to provide backup pacing, while non-selective left bundle branch pacing (ns-LBBP – V6RWPT: 68 ms, V1: rSR’ morphology) was successfully achieved with the ventricular lead (C315 sheath, MDT 3830 lead), resulting in ns-LBBP to left ventricular septal pacing (LVSP) transition (2V->1V) during threshold test (Figure 2, 3). During the post-procedural threshold test, transition from non-selective to selective LBBP was observed (nsLBBP->sLBBP) (Figure 4).

Later on the day of implantation, a 150 bpm incessant ventricular tachycardia (VT) occurred (Figure 5). Treatment with antiarrhythmic drugs, including beta-blocker, amiodarone and lidocaine provided only temporary rhythm control. Echocardiography revealed significant pericardial effusion causing hemodynamic instability. Pericardiocentesis was therefore performed, after which the chest X-ray demonstrated that the pericardial drain was in the projection of the right ventricle. The patient was urgently transported to the cardiac surgery department where following a short resuscitation, a pericardiotomy was performed. A large amount of blood and hematoma was evacuated and the puncture wound at the apex of the right ventricle was closed with a Teflon suture.
Following cardiac surgery, the patient required low-dose norepinephrine for circulatory support. In the early postoperative phase the incessant VT recurred and antiarrhythmic drugs including amiodarone again failed to provide adequate rhythm control. The longer the arrhythmia went on the more it compromised the patient’s hemodynamics. Considering the potential proarrhythmic effect of the implanted ventricular lead, ventricular overdrive pacing via a temporary wire was initiated and proved effective only for a limited period of time.

On January 18, 2023, electroanatomical mapping and focal radiofrequency (RF) ablation of the right ventricular anteroseptal region were performed, resulting only temporary suppression of the arrhythmia (Figure 6). The following day, the previously implanted conduction system pacing (CSP) ventricular lead was extracted, and a new conventional right ventricular lead was implanted in a high septal position (Figure 7, 8). Due to concerns regarding possible coronary artery damage caused by the initially implanted CSP lead, coronary angiography was performed. This revealed a mildly atherosclerotic coronary artery system without significant stenosis, as well as myocardial bridging of the proximal segment of LAD (Figure 9).

Because of persistent drug-refractory VT, a second focal RF ablation was performed on January 20, targeting the mid-septal region of the right ventricle, corresponding to the previous LBBAP lead position (Figure 10). However, the arrhythmia terminated during RF ablation and at the end of the procedure no ventricular arrhythmia was inducible, neither with high dose isoproterenol, after discontinuation of the sedation and extubation of the patient the index VT recurred incessantly with a frequency of 135 bpm.

Due to deteriorating hemodynamics a third ablation procedure had to be performed with extensive mid-septal applications from both the right and left ventricular sides (Figure 11). Ventricular arrhythmia was not inducible anv more and this time the patient remained arrhythmia-free on the long-term.
Control echocardiography showed heart failure with a reduced ejection fraction (LVEF: 30%), with diffuse hypokinesis and apical akinesis, which could be explained by the patient’s previous interventions. During the remainder of our observation period, no further arrhythmias occurred under antiarrhythmic therapy. Following a prolonged rehabilitation the patient was discharged.
During pacemaker follow-up, no arrhythmias were recorded by the device, while a high ventricular pacing burden (VP: 84%) was observed. In view of the heart failure with reduced ejection fraction, a cardiac resynchronization therapy (CRT) upgrade was recommended, which the patient declined while in a clinically stable condition.
With the dynamic development and increasing use of conduction system pacing, potential complications also has to be considered, including septal perforation, coronary artery injury, fistula formation, septal hematoma, and the development of ventricular arrhythmias. In the present case, we initially hypothesized that the ventricular arrhythmia was caused by local myocardial irritability related to the intramyocardial position of the CSP lead. However, as the ventricular tachycardia recurred after the LBBAP lead explantation, an alternative mechanism had to be considered: a proarrhythmic substrate resulting from local myocardial and/or microvascular injury caused by the initial CSP lead implantation.
In this specific case, multiple interventions – including repeated RF ablation procedures and explantation followed by reimplantation of the ventricular lead – were ultimately required to achieve successful termination of the drug-refractory incessant VT.