Patient History and Presentation
A 79-year-old woman with a dual-chamber pacemaker for sick sinus syndrome developed highly symptomatic atrial fibrillation (AF), subsequently regarded as permanent. Her previous treatment included cryoballoon ablation and antiarrhythmic therapy with flecainide followed by amiodarone, with AF recurrence and hyperthyroidism documented in the clinical history. She subsequently underwent repeat radiofrequency ablation targeting complex fractionated atrial electrograms, creation of a left atrial roof line, and isthmus ablation. A focal atrial tachycardia at the coronary sinus ostium was also ablated.
Severe fibrotic atrial cardiomyopathy, described as approximately 80% fibrosis, had been documented. A subsequent rate-control strategy was ineffective. Recurrent cardiac decompensation and tachycardia-mediated cardiomyopathy, with a left ventricular ejection fraction (LVEF) of 45% and NYHA class III–IV symptoms, prompted a planned upgrade to a cardiac resynchronisation therapy pacemaker (CRT-P) with atrioventricular (AV) node ablation.
The initial strategy was conventional biventricular CRT-P implantation. However, the coronary venous anatomy did not permit a suitable LV lead position: testing at the available sites resulted in phrenic nerve stimulation, high capture thresholds, or absent capture. Coronary venous LV lead placement was therefore unsuccessful (Figure 1). Therefore, strategy changed to conduction system pacing (CSP) as a bailout approach.

Figure 1. Unsuccessful coronary venous LV lead placement because of phrenic nerve stimulation, high capture thresholds, or absent capture at the available positions.

A lead was successfully implanted in the left bundle branch area, establishing left bundle branch area pacing (LBBAP). The lead position is shown fluoroscopically in an LAO 40° projection (Figure 2) and in an echocardiographic four-chamber view (Figure 3). The pace-and-ablate strategy was completed with AV node ablation.

Figure 2. Left bundle branch area lead position in an LAO 40° projection.
Figure 3. Four-chamber view showing the LBB lead position.

The surface ECG demonstrated a substantially narrower paced QRS complex during LBB pacing than during RV pacing (Figure 4). The recorded QRS durations were 80 ms and 120 ms, respectively (Figure 5).

Figure 4. Twelve-lead ECG comparison showing LBB pacing in the left portion and RV pacing in the right portion. The paced QRS complex is narrower during LBB pacing.
Figure 5. Annotated comparison and clinical outcomes: QRS duration 120 ms with RV pacing versus 80 ms with LBB pacing; BNP 2313 to 452 pg/mL; LVEF 45% to 61%; and follow-up NYHA class I–II.

Follow-Up
Following LBB pacing and AV node ablation, LVEF improved from 45% to 61%, BNP decreased from 2313 to 452 pg/mL, and functional status improved to NYHA class I–II (Figure 5). The most striking observation was the return of sinus rhythm despite the previous designation of permanent AF and extensive atrial disease. Follow-up was characterised by predominantly sinus rhythm. A representative device recording documents sinus rhythm at follow-up (Figure 6).

Figure 6. Representative device recording documenting sinus rhythm at follow-up.

Discussion
Arrhythmia Reduction With LBBAP
This case raises the possibility that physiological ventricular activation and haemodynamic recovery may favour atrial rhythm stability. In the observational I-CLAS analysis, 1414 propensity-matched CRT recipients with LVEF ≤35% were compared. Sustained VT/VF occurred less often with LBBAP than with biventricular pacing (4.2% versus 9.3%; HR 0.46, 95% CI 0.29–0.74). Among 890 patients without previous AF, new-onset AF was also less frequent (2.8% versus 6.6%; HR 0.34, 95% CI 0.16–0.73).[1] These findings support an association with lower arrhythmic risk.

Predictors of Sinus Rhythm Recovery
Direct evidence comes mainly from conventional CRT cohorts. Gasparini et al. reported spontaneous sinus rhythm resumption in 34 of 330 patients with permanent AF (10.3%), at a median of four months after CRT. Independent predictors were LV end-diastolic diameter ≤65 mm, left atrial diameter ≤50 mm, paced QRS duration ≤150 ms, and AV junction ablation.[2] These are observational associations, not validated selection criteria for withholding AV node ablation.
In a separate cohort of 65 CRT recipients with paroxysmal or persistent AF, Nedios et al. associated older age and larger left atrial diameter with progression to permanent AF, whereas greater biventricular pacing percentages were associated with stable sinus rhythm. Rhythm-control interventions were frequently used, and patients already classified as having permanent AF were excluded; this study therefore informs rhythm stability rather than spontaneous recovery from permanent AF.[3] Results are not uniform: Kiès et al. observed no spontaneous sinus rhythm restoration at six months despite reverse remodelling in a 74-patient chronic-AF cohort.[4]
In the present patient, advanced age and extensive reported atrial fibrosis make the rhythm outcome particularly noteworthy. Her narrow paced QRS is consistent with one favourable marker identified in the CRT literature, but does not establish why AF subsided. Ventricular recovery, rate regularisation after AV node ablation, prior atrial ablation, and concomitant treatment may all have contributed. The term permanent AF describes an accepted treatment strategy and does not necessarily imply biological irreversibility.[5]

Could Pace and Wait Be an Option
A staged approach—implanting CSP and reassessing rhythm before committing to AV node ablation—is a hypothesis for prospective study. Potential study candidates might be clinically stable patients with an independent pacing indication, adequate ventricular rate control, and a high effective CSP percentage without AV node ablation. Smaller atrial and ventricular dimensions could be explored as enrichment factors, but no validated LBBAP-specific prediction rule currently follows from these studies. The optimal observation period and criteria for proceeding to ablation also remain undefined.

Conclusion
LBB pacing combined with AV node ablation was followed by marked functional recovery and a return to predominantly sinus rhythm in a patient previously considered to have permanent AF. The case invites a focused question: can selected, stable patients benefit from CSP followed by reassessment before AV node ablation? Prospective studies must determine who, if anyone, can safely follow this approach.

References

  1. Herweg B, Sharma PS, Cano Ó, et al. Arrhythmic Risk in Biventricular Pacing Compared With Left Bundle Branch Area Pacing: Results From the I-CLAS Study. Circulation. 2024;149(5):379–390. doi: 10.1161/CIRCULATIONAHA.123.06746
  2. Gasparini M, Steinberg JS, Arshad A, et al. Resumption of sinus rhythm in patients with heart failure and permanent atrial fibrillation undergoing cardiac resynchronization therapy: a longitudinal observational study. Eur Heart J. 2010;31(8):976–983. doi: 10.1093/eurheartj/ehp572
  3. Nedios S, Doering M, Darma A, et al. Predictors of rhythm outcomes after cardiac resynchronization therapy in atrial fibrillation patients: When should we use an atrial lead? Clin Cardiol. 2021;44(2):210–217. doi: 10.1002/clc.23527
  4. Kiès P, Leclercq C, Bleeker GB, et al. Cardiac resynchronisation therapy in chronic atrial fibrillation: impact on left atrial size and reversal to sinus rhythm. Heart. 2006;92(4):490–494. doi: 10.1136/hrt.2005.064816
  5. Van Gelder IC, Rienstra M, Bunting KV, et al. 2024 ESC Guidelines for the management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS). Eur Heart J. 2024;45(36):3314–3414. doi: 10.1093/eurheartj/ehae176