Patient History and Presentation
A 67-year-old woman with highly symptomatic permanent atrial fibrillation was referred to our centre with progressive exercise intolerance, reduced left ventricular ejection fraction (LVEF), and NYHA class III symptoms.
Before referral, she had undergone three pulmonary vein ablation procedures and an additional ablation for right atrial tachycardia. Following further recurrences, the treating centre pursued a pace-and-ablate strategy, implanting a CRT-P followed by atrioventricular-node ablation.
Shortly after implantation, left ventricular (LV) pacing caused phrenic nerve stimulation. The patient found these symptoms highly distressing and insisted that the LV lead should be deactivated. Subsequent exclusive right ventricular (RV) pacing was associated with progressive functional deterioration and declining LVEF, raising concern for developing pacing-induced cardiomyopathy (PICM). The ECG during exclusive RV pacing is shown in Figure 1.

Figure 1. Exclusive right ventricular pacing after deactivation of the LV lead because of phrenic nerve stimulation. This pacing configuration was associated with subsequent clinical deterioration and reduced LVEF.

Treatment at Our Centre
We recommended restoring ventricular synchrony through left bundle branch area pacing (LBBAP). During the revision procedure, alternative pacing vectors of the existing quadripolar LV lead were also evaluated. Pacing using the proximal electrodes provided suitable LV stimulation without phrenic nerve stimulation.
However, the long-term usability of the LV lead remained uncertain. Furthermore, conventional biventricular (BiV) pacing produced a markedly prolonged QRS complex (Figure 2), suggesting suboptimal electrical resynchronisation. We therefore proceeded with replacement of the RV lead by a lead positioned in the left bundle branch area.

Figure 2. Conventional biventricular pacing. The 12-lead ECG shows a broad paced QRS complex with an annotated duration of 168 ms, consistent with suboptimal electrical resynchronisation in this patient.

The RV lead was successfully explanted, and a stylet-driven lead was implanted in the left bundle branch area. In addition to conventional electrocardiographic measurements, ultra-high-frequency electrocardiography (UHF-ECG) was used to assess lead positioning and compare ventricular activation during RV pacing, BiV pacing, left bundle branch pacing (LBBP) and LOT-CRT.
At the final lead position, the V6 R-wave peak time (V6-RWPT) was 78 ms, and the V6–V1 interpeak interval was 52 ms (Figure 3). UHF-ECG demonstrated a substantially more homogeneous ventricular activation pattern during LBBP than during conventional BiV pacing (Figure 4, left and middle panels).

Figure 3. Left bundle branch pacing at the final lead position. The annotated measurements show a V6 R-wave peak time of 78 ms and a V6–V1 interpeak interval of 52 ms. These measurements were assessed alongside QRS morphology and UHF-ECG findings.

Combined LBBP and LV pacing was subsequently optimised by adjusting the interventricular (VV) pacing interval, with LBBP preceding LV stimulation. This configuration also improved the delayed activation observed in the V8 region (Figure 4, right panel). Interpretation of UHF-ECG activation maps requires caution at large VV offsets; nevertheless, the optimised configuration suggested an additional improvement in electrical synchrony.

Figure 4. Surface ECGs and corresponding UHF-ECG activation maps during conventional biventricular pacing (BiV, left), left bundle branch pacing (LBB, middle), and combined LBB and LV pacing (LOT-CRT, right). Compared with BiV pacing, LBBP shows a more homogeneous activation pattern. Following optimisation of the VV interval with LBBP preceding LV stimulation, LOT-CRT also improves the residual activation delay in the V8 region. UHF-ECG map interpretation is limited at large VV offsets, and the maps should be assessed together with the conventional ECG and pacing settings.

Follow-Up
Following revision to LOT-CRT-P, the patient showed both clinical and echocardiographic improvement compared with her pre-revision status.

Conclusion
This case illustrates the potential role of LOT-CRT-P as an alternative when conventional BiV pacing provides suboptimal resynchronisation in a pace-and-ablate strategy. In this patient, LBBP improved ventricular activation, while the addition of appropriately timed LV pacing further addressed residual activation delay.
Alongside established electrocardiographic measurements, UHF-ECG may provide complementary information for comparing ventricular activation during different pacing configurations and guiding individual optimisation.